MI 3309 DeltaGT
Autotest sequences
61
6.5.5 Differential leakage measurement
The test is offered if it is applicable according to the autotest setting. The Differential
leakage current starting screen is displayed first. Measurement and options in Differential
leakage starting screen are described in chapter
5.2.8 Differential leakage
.
Figure 6.33: Differential leakage starting screen
After the measurement is carried out the Differential leakage result screen is displayed.
Figure 6.34: Differential leakage result screen
There are no special options to be set in the Differential leakage result screen.
6.5.6 Insulation resistance - P measurement
The test is offered if it is applicable according to the autotest setting. The Insulation
resistance - P starting screen is displayed first. Measurement and options in Insulation
resistance - P starting screen are described in chapter
5.2.4 Insulation resistance - P
.
Figure 6.35: Insulation resistance - P starting screen
After the measurement is carried out the Insulation resistance - P result screen is
displayed.
Figure 6.36: Insulation resistance - P result screen
Options in Insulation resistance - P result screen:
NEXT
Proceeds with the next measurement.
REPEAT
Repeats the test (use in case of multiple isolated / SELV / PELV
accessible points). Lowest result will be stored.