MI 3309 DeltaGT
Autotest sequences
54
6.1.1.7 Insulation resistance - P measurement
This test is offered if it is applicable according to the VDE Organizer setting.
The measurement is described in chapter
5.2.4 Insulation resistance - P
.
Figure 6.12: VDE Organizer autotest – Insulation resistance - P starting screen
Options in Insulation resistance-P starting screen:
NO
To be set if insulation test is not applicable
YES
To be set if insulation test is applicable
500 V
Standard test voltage
250 V
To be set if overvoltage protection devices are installed or
SELV/ PELV protection measures.
Figure 6.13: VDE Organizer – Insulation resistance - P result screen
Options in Insulation resistance - P result screen:
NEXT
Proceeds with the next measurement.
REPEAT
Repeats the test (use in case of multiple isolated / SELV / PELV
accessible points). Lowest result will be stored.
6.1.1.8 Selection of touch leakage current test method
The touch leakage test is offered if it is applicable according to the VDE Organizer setting.
If more test methods are possible the applicable touch leakage current test method must
be set first.
Options are
Touch leakage
and
Substitute leakage-P
.
Figure 6.14: Touch leakage current test method selection screen
6.1.1.9 Substitute leakage - P current measurement
The measurement is described in chapter
5.2.6 Substitute leakage - P.