MI 3309 DeltaGT
Technical specifications
18
3.3 Substitute leakage current, Substitute leakage – P current
Range Resolution Accuracy
0.00 mA
9.99 mA
0.01 mA
10.0 mA
20.0 mA
0.1 mA
(5 % of r 3 digits)
Powered by............................... battery or mains
Open circuit voltage .................. <50 V AC at rated mains voltage
Short circuit current................... <40 mA
Pass levels:
Substitute leakage .................... 0.25 mA, 0.50 mA, 0.75 mA, 1.00 mA, 1.50 mA, 2.00 mA,
2,25 mA, 2.50 mA, 3.50 mA, 4.00 mA, 4.50 mA, 5.00 mA,
5.50 mA, 6.00 mA, 7.00 mA, 8.00 mA, 9.00 mA, 10.0 mA,
15.0 mA, ---- mA
Substitute leakage - P............... 0.25 mA, 0.50 mA, 0.75 mA, 1.00 mA, 1.50 mA, 2,00 mA,
--- mA
Test duration............................. 2 s, 3 s, 5 s, 10 s, 30 s, 60 s, 120 s, --- s
Displayed current...................... calculated to appliance nominal mains supply voltage
(230 V)
Test terminals (Sub. leakage)
...
LN (Test socket)
↔
PE (Test socket)
LN (terminal)
↔
PE (terminal) (for fixed installed
appliances)
LN (Test socket)
↔
S/EB (probe)
LN (terminal)
↔
S/EB (probe) (for fixed installed
appliances)
Test terminals (Sub. leakage-P) LN (Test socket)
↔
S/EB (probe)
LN (terminal)
↔
S/EB (probe) (for fixed installed
appliances)
3.4 Polarity test
Powered by............................... battery or mains
Test voltage .............................. <50 V AC
Detects...................................... PASS, L OPEN, N OPEN, PE OPEN, L-N CROSS, L-N
SHORT, L-PE SHORT, N-PE SHORT, MULTIPLE FAULT.
Test terminals ........................... Test socket
↔
IEC test terminal