MI 3309 DeltaGT
Autotest sequences
52
6.1.1.3 Insulation resistance measurement
This test is offered if it is applicable according to the VDE Organizer setting.
The measurement is described in chapter
5.2.3 Insulation resistance
.
Figure 6.5: VDE Organizer – Insulation resistance starting screen
Options in insulation resistance starting screen:
NO
To be set if insulation test is not applicable
YES
To be set if insulation test is applicable
500 V
Standard test voltage
250 V
To be set if overvoltage protection devices are installed or
SELV/ PELV protection measures.
Figure 6.6: VDE Organizer – Insulation resistance result screen
There are no special options to be set in the insulation resistance result screen.
6.1.1.4 Selection of leakage current test method
The leakage test is offered if it is applicable according to the VDE Organizer setting.
If more test methods are possible the applicable leakage current test method must be set
first.
Options are
Differential leakage
and
Substitute leakage
.
Figure 6.7: Leakage current test method selection screen
6.1.1.5 Substitute leakage measurement
The measurement is described in chapter
5.2.5 Substitute leakage
.
Figure 6.8: VDE Organizer – Substitute leakage starting screen
Options if
Appliances with heating elements
is set:
/
Sets power of heating elements.