MI 3309 DeltaGT
Autotest sequences
56
Figure 6.19: VDE Organizer –RCD starting screen
Options in RCD starting screen:
/
Sets nominal RCD value
Note:
The test mode is automatically set to Single test, 1xI
Δ
N
, both polarities.
Figure 6.20:
VDE Organizer –RCD result screen
There are no special options in the RCD result screen.
6.1.1.12 Functional test
Figure 6.21: VDE Organizer – Functional test result screens
Options in Functional test result screen:
PASS / FAIL
To be applied manually.
Option if
General
or
Appliance with heating elements
is set:
POWER
Starts Power test. The measurement is described in chapter
5.2.11
Power test
.
Option if
IEC leads, multiple outlets without electronic parts
is set:
POLARITY
Starts Polarity test. The measurement is described in chapter
5.2.7
Polarity test
Note:
After Power or Polarity test is completed, ESC key should be pressed to return to
Functional test result screen.