MI 3309 DeltaGT
Appendix B – Pre-programmed autotests (GER)
74
Appendix B – Pre-programmed autotests (GER)
Pre-programmed autotest sequences table
Autotest shortcut code
A01 A02 A03 A04
Kl_1_Iso Kl_1_Iso_BLT Kl_1_Ia
Kl_1_Ia_BLT
Visual test
Output
200 mA
200 mA
200 mA
200 mA
Limit
0.30
0.30
0.30
0.30
Earth
continuity
Time
5 s
5 s
5 s
5 s
Output
500 V
500 V
Limit
1.00 M
1.00
M
Insulation
Time
5 s
5 s
Output
500 V
Limit
2.00 M
Insulation
(probe)
Time
5 s
Output
40 V
40 V
Limit
3.50 mA
3.50 mA
Sub. leakage
Time
5 s
5 s
Output
40 V
Limit
0.50 mA
Sub. leakage
(probe)
Time
5 s
Output
230 V
230 V
Limit
3.50 mA
3.50 mA
Differential
leakage
Time
180 s
180 s
Output
230 V
Limit
0.50 mA
Touch
leakage
Time
180 s
Functional test
Output
230 V
230 V
230 V
230 V
Limit
Power*
Time
180 s
180 s
180 s
180 s
Output
Limit
TRMS clamp
current
Time
Polarity test
* optional