MI 3309 DeltaGT
Single test
47
Figure 5.38: Power test menu
Test parameters for the Power test
OUTPUT
Test voltage [MAINS voltage]
TIME
Measuring time
[2 s, 3 s, 5 s, 10 s, 30 s, 60 s, 120 s, --- s (continuous
measurement)]
Test circuit for the power test
Figure 5.39: Power test
Functional test procedure
Select the POWER function.
Set test parameters.
Connect device under test to the instrument’s test socket and switch it on (see
figure 5.39).
Connect the instrument to mains voltage.
Press the TEST key for measurement. To stop continuous measurement press
TEST key once again.
Store the result by pressing MEM key (optional).
Figure 5.40: Example of apparent power measurement result
Displayed results:
Main result ............. Apparent power