Single test
34
This function is primarily intended for testing Class II appliances and Class II parts of Class
I appliances.
Figure 5.12: Insulation resistance - P menu
Test parameters for Insulation resistance - P measurement
OUTPUT
Test
voltage
[250 V, 500 V]
LIMIT
Minimum resistance
[0.01 M
, 0.10 M
, 0,25M
, 0.30 M
, 0.50 M
,
1 M
, 2 M
, 4 M
, 7 M
, 10 M
, --- M
]
TIME
Measuring time
[2 s, 3 s, 5 s, 10 s, 30 s, 60 s, 120 s, --- s (continuous
measurement)]
Test circuits for Insulation resistance - P measurement
Figure 5.13: Measurement of Insulation resistance - P
Figure 5.14: Measurement of insulation resistance of fixed installed DUTs