8
TEMPERATURE MANAGEMENT SYSTEM
O P E R A T O R ’ S M A N U A L
Inspect Fluid Delivery Line
1. Power On the system
2. From the Patient Therapy Selection screen press the
Hypothermia
button
to display the
Hypothermia
therapy screen.
3. From the
Hypothermia
therapy screen, press the
Manual Control
button
to open the
Manual Control
window.
4. Set the
Manual Control
water target temperature to 28°C and the
duration to 30 minutes.
5. Connect a shunt to a set of fluid delivery line ports.
6. Press the
Help
button and then press the
Help Index
button. Select the
topic
Maintenance and Service
and sub topic
System Diagnostics
then
press the
Display
button. Verify that inlet pressure is -7 ± 0.2.
7. Repeat on all valves. If inlet pressure is out of range, replace the two
valves that the shunt is connected to.
8. Ensure that the shunt is removed before device is put back in service.
Software Update
Software updates will be provided on a flash drive. Installation of software
updates will be performed via the USB port on the front of the control module.
The software update feature will automatically initiate if the control module detects
the proper files on a flash drive inserted in the USB port at power On.
To install software update:
1) Insert flash drive into the USB port.
2) An image of a timer will display while the software update in being installed,
and will disappear when the software installation process is completed.
3) After installation, the new software version will display in the
Software
Version
field in
Advanced Setup
.
Service
Contact Customer Support for technical support and customer service
instructions to enable appropriately qualified technical personnel to repair those
parts of the equipment that Medivance considers repairable.
Calibration
See
Arctic Sun
®
Temperature Management System Service Manual for
calibration requirements and instructions. Calibration is recommended after
2,000 hours of operation or 250 uses, whichever occurs first.
Appendix A: Product Specifications
Technical Description
The
Arctic Sun
®
Temperature Management System is a thermoregulatory
device that monitors and controls patient temperature within a range of 32°C
to 38.5°C (89.6°F to 101.3°F). The Arctic Sun
®
Temperature Management
System consists of the Control Module and disposable ArcticGel
™
Pads.
A patient temperature probe connected to the Control Module provides
patient temperature feedback to an internal control algorithm which
automatically increases or decreases the circulating water temperature to
achieve a pre-set patient target temperature determined by the clinician.
The
Arctic Sun
®
Temperature Management System pulls temperature-
controlled water ranging between 4°C and 42°C (39.2°F and 107.6°F)
through the
ArcticGel
™
Pads at approximately 0.7 liter per minute per pad.
This results in heat exchange between the water and the patient.
The
Arctic Sun
®
Temperature Management System Control Module is
a CLASS I mobile device (Type BF, IPX0 and Mode of Operation –
Continuous) per classification scheme of IEC 60601-1.
The
Arctic Sun
®
Temperature Management System Control Module meets
both the electromagnetic interference and susceptibility requirements of IEC
60601-1, and is compatible with other equipment that also conforms to that
standard. There is no known failure mode in the
Arctic Sun
®
Temperature
Management System Control Module associated with electromagnetic
interference from other devices. See the
Arctic Sun
®
Temperature
Management System Service Manual for the full declaration regarding
electromagnetic compatibility.
Environmental Conditions
Temperature Range
Operating: ...................
10°C to 27°C (50°F to 80°F)
Storage: .................
-30°C to 50°C (-20°F to 120°F)
At operating temperatures higher than 27°C (80°F), the refrigeration system’s
cooling capacity and therefore its ability to cool a patient
is compromised.
Humidity Range (relative humidity, non-condensing)
Operating: .............................................. 5% to 70%
Storage: ................................................. 5% to 95%
Atmospheric Pressure Range:... 60 kPa to 110 kPa
Disposal
Upon end of life, dispose of in accordance with local WEEE regulations or
contact your local
Bard
®
Supplier or Distributor to arrange for disposal.
Summary of Contents for ARCTIC SUN Simply Advanced 5000
Page 1: ...TEMPERATURE MANAGEMENT SYSTEM SimplyAdvanced M O D E L 5 0 0 0 O P E R A T O R S M A N U A L ...
Page 13: ...13 E N G L I S H ...
Page 25: ...25 F R A N Ç A I S F R E N C H ...
Page 121: ...121 D A N S K D A N I S H ...
Page 133: ...133 S V E N S K A S W E D I S H ...
Page 145: ...145 S U O M I F I N N I S H ...
Page 157: ...157 N O R S K N O R W E G I A N 157 ...
Page 193: ...193 193 Č E S K Y C z e c h ...
Page 205: ...205 T Ü R K Ç E T u r k i s h ...
Page 217: ...217 R O M Â N Ă R o m a n i a n ...
Page 229: ...229 S L O V E N Č I N A S l o v a k i a ...
Page 273: ...273 한 국 어 K O R E A N ...