Technical data
15
Subject to technical changes.
200114
BA-42820-02-V03
Water sampling station
EASYPRO
Operating instructions
5.3.3 Membrane covered measuring cells for measuring chlorine dioxide
Measuring cell
CD 4.2
CD 4.2
Measuring input
Chlorine dioxide
Measuring range
0 – 0,5 mg/l ClO
2
0 – 2 mg/l ClO
2
Transfer/slope
4 – 20 mA
Electrolyte
Diaphragm cap with electrolyte
Connection
Free wire end
Pressure range
0 – 0,5 bar
Flow rate
30 l/h
Measuring principle
amperometric diaphragm covered cell measuring cell
Tab. 14: Membrane covered measuring cells for measuring chlorine dioxide
5.3.4 Diaphragm covered measuring cell for measuring total chlorine
Measuring cell
GCM
GCM
GCM
Measuring input
Total chlorine
Measuring range
0 – 2 mg/l Cl
ges.
0 – 5 mg/l Cl
ges.
0 – 10 mg/l Cl
ges.
Transfer/slope
4 – 20 mA
Electrolyte
Diaphragm cap with electrolyte
Connection
Free wire end
Pressure range
0 – 1 bar
Flow rate
30 l/h
Measuring principle
amperometric diaphragm covered cell measuring cell
Tab. 15: Diaphragm covered measuring cell for measuring total chlorine
5.3.5 Conductivity measurement (conductive)
Description
Value
Measuring range
mS/cm
max. 100 (5 % salt content), depending on input module
C factor
1
Electrical connection
M12x1 plug (A-coded)
Pressure range
bar
0 – 6
Measuring principle
Conductivity measurement (conductive)
temperature measurement
Integrated Pt100
Hydraulic connection
PG 13,5
Installation dimension
mm
120 x 12
Tab. 16: Technical data conductivity sensor