Fault resolution
49
Subject to technical changes.
200114
BA-42820-02-V03
Water sampling station
EASYPRO
Operating instructions
12 Fault resolution
See below for information about how to rectify faults on the control product or the system. If you cannot eliminate the fault, please consult with the man-
ufacturer on further measures or return the product for repair.
Fault
Possible causes
Measures for resolution
Balls in the disinfection
measuring cell rotate too
weakly or float does not
float
Water inflow pressure too low
n
Lay sample water line with larger cross-section.
n
Installation of a sample water pump.
Sample water filter soiled
Clean or replace the filter sieve.
The float floats but the reed
switch does not switch upon
connection.
Harmonisation between the float and reed
switch is incorrect.
Calibrate reed switch
Reed contact defective
Replace reed contact
Display fluctuates during
the excess chlorine
measurement of the
connected measurement
amplifier
Fluctuating pH value influences the chlorine
measurement
Stabilise the pH value in the water
(e.g. optimise control path, controller parameters)
“Sample water shortage”
display on the measurement
device.
Reed switch in the flow control defective
Replace reed switch
Inflow tap closed
Open the inflow tap (Fig. 2 „Water sampling station EASYPRO 4“ on
page 10, pos. 5).
The sample water inflow is soiled.
Check the visible parts and filter (Fig. 2 „Water sampling station
EASYPRO 4“ on page 10, pos. 3) for soiling and clean if necessary.
The LEDs are continually
red.
The controller does not control the LEDs
correctly.
Comply with the instructions in section8.3.2.6 „Actuation of the LEDs
for the water sampling stations“ on page 34.
Measured value fluctuates
constantly
Loose measurement cable connections
Check measurement cable; re-attach if necessary
The measurement is very
slow.
Diaphragm blocked
Clean electrode
Electrode becoming obsolete
Replace electrode
Buffer solution becoming obsolete
Check the expiration date and replace if necessary.
The measurement deviates
or is unstable
Measuring cell defective
Replace measuring cell
Diaphragm is dry and/or soiled
Clean measuring cell
The display deviates by a
constant value
Measuring cell not correctly installed
Check correct installation; check correct calibration
External potential in the system
Remove the external potential or set the offset (see chapter 9.5.2.3
„Offset compensation“ on page 38).
TOPAX
®
MC
controller
constantly displays pH value
"7.00".
Short circuit
Check measuring cell for damage and replace if necessary
No reaction to pH buffer
solution
Measuring cell not correctly installed
Check correct installation; check correct calibration
Measuring cell defective
Check measuring cell for damage and replace if necessary
Value for "free chlorine"
fluctuates heavily
pH-value instability
Check pH value measurement
Tab. 43: Troubleshooting