Fault resolution
50
BA-42820-02-V03
Water sampling station
EASYPRO
Operating instructions
Fault
Possible causes
Measures for resolution
The chlorine measuring cell
displays too much with
respect to the DPD-1
measurement
May occur when using "water fining
agents". Chlorine-oxygen bonds are
contained in some of these agents which
react with chlorine to make chlorine dioxide,
thus distorting the measuring signal by
greater values
Refrain from dosing this agent. A change of water might be required.
The DPD-1 method displays
excessive chlorine content
to which the measuring cell
cannot be adjusted
Use of "organic chlorine". When using
organic chlorine products, cyanuric acid
accumulates in the swimming pool water.
Replace the water completely and use other chemicals for disinfection
The
TOPAX
®
MC
controller
loses all settings after he
has been disconnected from
the network and then recon-
nected.
The battery is empty.
n
Check that the battery is really empty. To do so, navigate to the
menu System > Information > System values. You can view the
battery voltage under “Battery”. If the voltage is under 2.9 V,
change the battery.
n
Replace the battery. See section 11.8 „Battery“ on page 47.
The
TOPAX
®
MC
controller
is switched off.
The power supply has been interrupted.
Reconnect the current supply
The fuse of the
TOPAX
®
MC
controller is
damaged.
Replace the fuse. See section 11.9 „Replacing the fuse“ on page 47.
The sensor error is
displayed as an alarm.
The sensor has not been installed correctly.
Make sure that the sensor has been connected correctly. See section
7.7 „Connecting sensors“ on page 23.
The signal cable to the sensor has a break.
Replace the signal cable.
Tab. 43: Troubleshooting
EASYPRO