LumaSMART
®
User Manual
Installation of the LumaSMART® Controller
•
51
The Fault conditions include:
System Condition
Cause
Fot1PortOpenError
Communications with FOT 1
Fot2PortOpenError
Communications with FOT 2
Fot3PortOpenError
Communications with FOT 3
Fot4PortOpenError
Communications with FOT 4
ProbeError
Probe Signal Strength
BatteryError
Backup Clock Battery need replacement
DataLoggingError
File error with the Data Log
AsciiPortOpenError
Fault with the default RS232 Port
SerialPortOpenError
Fault with the secondary RS232/485 Port
EthernetPortError
Fault on the Ethernet Port
StartupDateError
System Clock is in error
RebootInProgress
Shutdown/Reboot is in progress
ProtocolWriteDataError
Fault within selected Protocol
RelayTestWasRun
Relay Testing
A date/time stamped log is generated when the System Status Relay changes states. See Section 5.2.6.5 for
4.3.4
Optional Communication Wiring
The Ethernet connector is the normal standard. Shielded Ethernet cable must be used. By default, the IP address
for the Ethernet connection is set to a static non-routable, local host, loopback IP address of 127.0.0.1. Before the
system can communicate to the outside world, the static IP address must be set to match your network
topology,see Section 5.2.6.7 for
Setup TCP/IP Ethernet Interface Parameters
LumaSMART Communication Connections
The Txd-/Txd+ pins are used for RS485 communications (2-wire/Half-Duplex). See Section 5.2.6.7 for
and
Setup RS485 Interface Parameters
. The Rx/Tx pins are RS232 and are used for the dedicated ASCII
Note:
Always connect the GND pin between the LumaSMART and the report communications
device.
Communication Wiring Diagram
Use the MC 1.5/8-SFT-3.81 connector from Phoenix Contact (Part # 1827732) to make the external connection to
the communication port. The optional ferrule is Part # 3200519.