LumaSMART
®
User Manual
User Commands
•
142
10.2
rtrv Command
The command
rtrv=X
retrieves X number of days of Data Log reports the LumaSMART. If the memory is empty, the
Data Manager replies Okay, and no data reports follow. If the requested number of data reports is greater than the
number of reports in memory, the Data Manager sends the available number of reports.
DATE :5/3/2013
TIME :00:00:44 P.E. 39.48 38.97 38.40#01000110000001001
TIME :00:01:44 P.E. 39.47 38.96 38.42#01000110000001001
TIME :00:02:44 P.E. 39.43 38.95 38.40#01000110000001001
TIME :00:03:44 P.E. 39.44 38.94 38.40#01000110000001001
TIME :00:04:44 DSBL 39.45 38.94 38.39#11000110000001001
TIME :00:05:44 DSBL 39.44 38.94 38.37#11000110000001001
TIME :00:06:44 DSBL 39.46 38.99 38.39#11000110000001001
TIME :00:07:44 DSBL 39.52 38.97 38.37#11000110000001001
TIME :00:08:44 DSBL 39.49 38.98 38.39#11000110000001001
TIME :00:09:44 DSBL 39.47 38.97 38.35#11000110000001001
TIME :00:10:45 DSBL 39.49 38.98 38.40#11000110000001001
This report shows the following information:
•
Date is May 3d 2013, Time range is 00:00:44 to 00:10:44
•
Channel 1 was in Probe Error, and then Disabled.
•
Channels 2, 3 and 4 were reading in the 38-39 °C range
•
Date is May 3d 2013, Time range is 00:00:44 to 00:10:44
•
Channel 1 was in Probe Error, and then Disabled.
•
Channels 2, 3 and 4 were reading in the 38-39 °C range
•
The System Status relay (1
st
digit after the #) showed system fault (Deenergized) until the channel with
the P.E. was disabled. It then when into a Normal “OK” state as shown by being Energized ‘1’.
•
Relays 1, 5, 6, 13 and 16 are Energized ‘1’
•
Relays 2, 3, 4, 7, 8, 9, 10, 11, 12, 13 and 15 are Deenergized ‘0’