Instructions for Use Supplement B – CO
2
Sensor LeoCap
|
Rev. 04
23
Operating conditions
Ambient temperature 10 – 35 °C
Relative humidity
10 – 90% (at ambient temperature of 25 °C)
Storage conditions
Ambient temperature 5 – 40 °C
Relative humidity
<80 % (at ambient temperature of 25 °C)
Atmospheric pres-
sure
390 – 900 mmHg
Transport conditions
Ambient temperature -50 – 50 °C
Relative humidity
<80 % (at ambient temperature of 25 °C)
Standards compliance
•
Safety: the device complies with IEC 60601-1, ISO 80601-2-55.
•
Electromagnetic compatibility (EMC): the device meets the requirements
of IEC 60601-1-2. The device is intended for use in the electromagnetic
environment specified in chapter 10.
Remark
The following factors may impair the CO
2
measuring accuracy:
• mechanical damage of the device
• cyclical pressure peaks of up to 10 kPa (100 cmH
2
O)
• high gas humidity or condensate
• leakage or internal venting of the sampled gas
Remark
The respiratory rate (RR) has no influence on the measuring accu
-
racy of the CO
2
concentration within the specified RR measuring
range and at an I:E ratio of 1:1.
Remark
There is no drift in the CO
2
measuring accuracy for at least 6
hours.
Summary of Contents for elisa 300
Page 3: ...elisa 300 elisa 500 Instructions for Use Basic Manual ...
Page 62: ...5 General information 60 Basic Instructions for Use elisa 300 500 SW 2 09 x Rev 11 ...
Page 217: ...elisa 300 elisa 500 Instructions for Use Supplement A Manoeuvres Functions Ventilation Modes ...
Page 351: ...CO2 Mainstream Sensor LeoCap Instructions for Use Supplement B ...
Page 379: ...Multi Gas Sensor LeoLyzer Instructions for Use Supplement C ...
Page 407: ...Pulse Oximetry and LeoClac Instructions for Use Supplement D ...
Page 437: ...Medical Equipment Interface elisa megs Instructions for Use Supplement F ...
Page 440: ...4 Instructions for Use Supplement F elisa megs Rev 03 This page is intentionally left blank ...