A 3 | Functions
80
Instructions for Use elisa 300-500 | Supplement A |
SW 2.09.x
| Rev. 11
A 3.14
Bronchoscopy
You reach the “Bronchoscopy” function by pressing the “System” button and then
by selecting the “Extras” tab.
During bronchoscopy and similar therapeutic measures, which are associated
with high leakage, frequent audible alarms cannot be prevented. To reduce the
stress level during the intervention, audible alarms are often permanently muted.
Activation of the bronchoscopy function will automatically silence the audible
ventilation alarms. In this situation, it is essential to watch the patient closely or
use other suitable means of patient monitoring.
Remark
When the bronchoscopy function is activated, you will see the
“Bronchoscopy” label in the display area of the ventilation mode
concerned.
Warning
Activating bronchoscopy will disable all audible and visual ventila-
tion alarms. Technical alarms, however, are still reported.
Warning
To rule out any risk for the patient, the bronchoscopy function
may only be activated when close patient monitoring is ensured.
The extent of direct observation of the patient or additional mon
-
itoring methods (pulse oximetry, end-tidal CO
2
measurement, ...)
depend on the patient’s clinical condition.
Warning
After deactivation of the bronchoscopy function, the default alarm
limits will automatically be restored. To rule out any risk for the
patient, adapt the alarm limits of the device to the alarm limits of
the therapeutic measures that you need to take for the patient at
hand.
Remark
After deactivation of the bronchoscopy function, an audible alarm
will be emitted to remind the user of the necessity to adapt the
alarm limits. When adjusting the first alarm limit, the alarm criteri
-
on will be deactivated.
Summary of Contents for elisa 300
Page 3: ...elisa 300 elisa 500 Instructions for Use Basic Manual ...
Page 62: ...5 General information 60 Basic Instructions for Use elisa 300 500 SW 2 09 x Rev 11 ...
Page 217: ...elisa 300 elisa 500 Instructions for Use Supplement A Manoeuvres Functions Ventilation Modes ...
Page 351: ...CO2 Mainstream Sensor LeoCap Instructions for Use Supplement B ...
Page 379: ...Multi Gas Sensor LeoLyzer Instructions for Use Supplement C ...
Page 407: ...Pulse Oximetry and LeoClac Instructions for Use Supplement D ...
Page 437: ...Medical Equipment Interface elisa megs Instructions for Use Supplement F ...
Page 440: ...4 Instructions for Use Supplement F elisa megs Rev 03 This page is intentionally left blank ...