Operation | 7
Basic Instructions for Use elisa 300-500
|
SW 2.09.x
|
Rev. 11
77
Depending on the configured option, the compensation applies only to the spon
-
taneous breaths or also to the pressure-controlled mandatory breaths.
Tube compensation for pressure-con-
trolled and spontaneous breaths
Tube compensation only for spontaneous
breaths
Remark
In the ventilation of neonates and infants, the compensation of
the tube-related work of breathing activated for the mandatory
breaths may increase the tidal volume.
Set the desired degree of compensation in percent and the internal diameter of
the tube in millimetres.
Warning
The tube diameter must always be set to the size of the tube used.
A tube diameter that is too small can lead to overcompensation
and oscillation during ventilation.
Remark
The internal diameter is printed on the outside of the tube or its
packaging.
Inspiratory or expiratory tube compensation can be activated once the degree of
compensation and tube diameter have been set.
When the tube compensation is activated, the device calculates the tracheal pres-
sure at the Y-piece before the tube and the pressure is displayed by an additional
curve. If the curve is not visible, it may have been deactivated in the configuration
menu.
Remark
If the compensation values are high, the activation of expiratory
compensation can lead to auto-triggering of the pressure support.
In this case, a higher trigger sensitivity should be selected.
Remark
Visible overshoots and undershoots are not transferred to the
patient via the tube.
Summary of Contents for elisa 300
Page 3: ...elisa 300 elisa 500 Instructions for Use Basic Manual ...
Page 62: ...5 General information 60 Basic Instructions for Use elisa 300 500 SW 2 09 x Rev 11 ...
Page 217: ...elisa 300 elisa 500 Instructions for Use Supplement A Manoeuvres Functions Ventilation Modes ...
Page 351: ...CO2 Mainstream Sensor LeoCap Instructions for Use Supplement B ...
Page 379: ...Multi Gas Sensor LeoLyzer Instructions for Use Supplement C ...
Page 407: ...Pulse Oximetry and LeoClac Instructions for Use Supplement D ...
Page 437: ...Medical Equipment Interface elisa megs Instructions for Use Supplement F ...
Page 440: ...4 Instructions for Use Supplement F elisa megs Rev 03 This page is intentionally left blank ...