Functions | A 3
Instructions for Use elisa 300-500 | Supplement A |
SW 2.09.x
| Rev. 11
35
Remark
Check that the Cuffscout line fits the Paux I port. If it doesn’t,
your device was probably not yet upgraded to the new Paux ports.
Please contact your service technician.
A 3.4.1.1
Cuffscout line
Luer male
cuff
elisa 500
Slip the end with the non-return valve onto the Paux I port of elisa 500.
The valve allows gas to flow through in only one direction. Therefore, the cuff
pressure is maintained even if the Cuffscout line becomes disconnected.