A 3 | Functions
64
Instructions for Use elisa 300-500 | Supplement A |
SW 2.09.x
| Rev. 11
Remark
Multi-gas measurement during NIV:
During non-invasive ventilation, leakage of varying degrees and de-
pendent on the ventilation pressure occurs regularly. This means
that the volume expired by the patient and the measured expira
-
tory volume will not be the same, which has a considerable effect
on the accuracy of the measurement. Therefore, it is necessary to
critically review the measured values and to evaluate them in the
clinical context.
Remark
The multi-gas sensor LeoLyzer must be configured on the con
-
figuration level of elisa 300 / elisa 500 in order to be available
for multi-gas measurement. You can review the current interface
configuration on the standby or ventilation screen under the “Sta
-
tus” tab by selecting the second view with a horizontal swipe. If in
doubt, contact your service technician.
A 3.11.2
Measuring principle
The method is based on the measurement of infrared light absorption at wave-
lengths of 4.2 μm, 7.85 μm, and 8.3 μm. The measured amount of light that passes
through the gas to the photo sensor serves as the calculation basis. The con-
centration of CO
2
and anaesthetic gas is calculated from the partial pressure at
atmospheric pressure conditions.
A 3.11.3
Power supply
The multi-gas sensor is powered via the serial (BF) interface of the intensive care
ventilator.
If the main power supply fails, the sensor continues to be powered from the inten-
sive care ventilator as long as the battery of the intensive care ventilator maintains
the ventilation function.
A 3.11.4
Breathing circuit
The CO
2
sensor is compatible with all approved breathing circuits.
A 3.11.5
Integrating the multi-gas sensor LeoLyzer
Insert the airway adapter between the Y-piece and the tube of the breathing cir-
cuit.
Connect the sampling line to the airway adapter by turning the line clockwise.
Connect the sampling line to the water trap by turning the line clockwise.
Summary of Contents for elisa 300
Page 3: ...elisa 300 elisa 500 Instructions for Use Basic Manual ...
Page 62: ...5 General information 60 Basic Instructions for Use elisa 300 500 SW 2 09 x Rev 11 ...
Page 217: ...elisa 300 elisa 500 Instructions for Use Supplement A Manoeuvres Functions Ventilation Modes ...
Page 351: ...CO2 Mainstream Sensor LeoCap Instructions for Use Supplement B ...
Page 379: ...Multi Gas Sensor LeoLyzer Instructions for Use Supplement C ...
Page 407: ...Pulse Oximetry and LeoClac Instructions for Use Supplement D ...
Page 437: ...Medical Equipment Interface elisa megs Instructions for Use Supplement F ...
Page 440: ...4 Instructions for Use Supplement F elisa megs Rev 03 This page is intentionally left blank ...