Manoeuvres | A 2
Instructions for Use elisa 300-500 | Supplement A |
SW 2.09.x
| Rev. 11
19
A 2.6
PEEPfinder manoeuvre
With the PEEPfinder, the device offers a respiratory mechanics manoeuvre during
which a quasi-static P/V graph (pressure/volume) is recorded at the bedside and
which assists in determining a secure window for the ventilation settings. The
PEEPfinder can optionally be used as a recruitment tool.
Remark
The extent of the manoeuvre (only inspiratory measurement, only
expiratory measurement, or inspiratory and expiratory measure
-
ment) can be set on the configuration level.
A pre-oxygenation function with an adjustable oxygen concentration can be acti
-
vated to prevent inadequate oxygen supply during the manoeuvre.
Depending on the configuration, the manoeuvre covers only the inspiratory por
-
tion, only the expiratory portion, or the entire breath.
The measured data may prove very useful for the clinical diagnosis as well as for
optimizing the ventilation settings.
A 2.6.1
Prerequisites
Conditions
for using PEEPfinders are:
• The patient must be intubated or tracheotomized and ventilated.
• The device passed the system test successfully (including measurement of the
breathing circuit).
• Only mandatory ventilation is applied, the patient is not breathing spontane-
ously. In spontaneous breathing modes, during HFOT or non-invasive ventila-
tion, the PEEPfinder function is deactivated.
• The overall leakage is low.
A 2.6.2
Contraindications
Possible
contraindications
for using PEEPfinder:
•
Patients breathing spontaneously or with phases of spontaneous respiration
•
Patients whose cardiovascular condition is unstable
• Hypovolaemic patients
•
Patients with suspected or confirmed intracranial pressure problems
•
Patients unable to tolerate elevated intrapulmonary pressures for other rea
-
sons
•
Patients unable to tolerate extended ventilation pauses
•
Patients susceptible to barotrauma or volutrauma
•
Patients with an inadequate depth of sedation and level of relaxation
Summary of Contents for elisa 300
Page 3: ...elisa 300 elisa 500 Instructions for Use Basic Manual ...
Page 62: ...5 General information 60 Basic Instructions for Use elisa 300 500 SW 2 09 x Rev 11 ...
Page 217: ...elisa 300 elisa 500 Instructions for Use Supplement A Manoeuvres Functions Ventilation Modes ...
Page 351: ...CO2 Mainstream Sensor LeoCap Instructions for Use Supplement B ...
Page 379: ...Multi Gas Sensor LeoLyzer Instructions for Use Supplement C ...
Page 407: ...Pulse Oximetry and LeoClac Instructions for Use Supplement D ...
Page 437: ...Medical Equipment Interface elisa megs Instructions for Use Supplement F ...
Page 440: ...4 Instructions for Use Supplement F elisa megs Rev 03 This page is intentionally left blank ...