10 | Cleaning, disinfection and sterilisation, disposal
174
Basic Instructions for Use elisa 300-500
|
SW 2.09.x
|
Rev. 11
Remark
Use a cleaning agent suitable for anaesthesia accessories.
For machine cleaning with disinfection, “neodisher MediClean forte” by “Chemis-
che Fabrik Dr. Weigert GmbH & Co. KG” and Schülke thermosept® X.tra, applied
according to the manufacturer's instructions, is recommended.
Select a programme for anaesthesia accessories (e.g. Vario TD). Cleaning is per
-
formed at 55°C (131°F) for a minimum of 10 minutes and thermal disinfection at
93°C (199.4°F) for a minimum of 5 minutes.
The final rinse must be carried out with demineralised water.
After cleaning and disinfection take the components of the valve bar out of the
washer disinfector and carry out a visual inspection.
Subsequently, allow the parts to dry for at least 40 minutes.
10.2.2.3
Sterilisation of the valve bar
The valve bar can be steam sterilised at 134°C (273.2°F).
Minimum sterilisation period is 3 minutes.
Alternative: The A
0
value must be at least 3000.
Use DIN EN 285 compliant vacuum steam sterilisers for sterilisation, preferably
with fractional vacuum.
Remark
Observe the instructions for use of the autoclave.
Remark
Autoclave the expiratory flow sensor, expiratory and inspiratory
valves assembled.
Warning
Ethylene oxide diffusing into the components of the valve bar
pose a health hazard to the patient. Therefore, do not sterilise the
valve bar with ethylene oxide.
Warning
All silicone parts of the valve bar must be replaced after 500 ster-
ilisation cycles or during annual maintenance, whichever occurs
first.
Summary of Contents for elisa 300
Page 3: ...elisa 300 elisa 500 Instructions for Use Basic Manual ...
Page 62: ...5 General information 60 Basic Instructions for Use elisa 300 500 SW 2 09 x Rev 11 ...
Page 217: ...elisa 300 elisa 500 Instructions for Use Supplement A Manoeuvres Functions Ventilation Modes ...
Page 351: ...CO2 Mainstream Sensor LeoCap Instructions for Use Supplement B ...
Page 379: ...Multi Gas Sensor LeoLyzer Instructions for Use Supplement C ...
Page 407: ...Pulse Oximetry and LeoClac Instructions for Use Supplement D ...
Page 437: ...Medical Equipment Interface elisa megs Instructions for Use Supplement F ...
Page 440: ...4 Instructions for Use Supplement F elisa megs Rev 03 This page is intentionally left blank ...