Functions | A 3
Instructions for Use elisa 300-500 | Supplement A |
SW 2.09.x
| Rev. 11
57
A 3.10.8
Connecting the sidestream sensor ISA CO
2
Back
CO
2
adapter
Connect the CO
2
sensor ISA CO
2
to the configured BF interface on the back
of the ventilator using the connection cable and the CO
2
adapter (see separate
accessory list).
Using the approved support, fasten the sidestream sensor to one of the two side
handles of the ventilator. Using a sampling line, connect the sensor to the appro-
priate port of the HME filter.
Warning
The sampling line must be replaced after each patient to avoid
cross-contamination.
When using a sidestream CO
2
sensor, you can choose between sampling lines
with Luer-Lock connectors (such as on the HME filter) or with a tube adapter
(e.g., for combination with heated tubing systems). The respective sampling line
is first connected to the gas inlet of the sidestream CO
2
sensor and then to the
corresponding connection in the breathing circuit.
Summary of Contents for elisa 300
Page 3: ...elisa 300 elisa 500 Instructions for Use Basic Manual ...
Page 62: ...5 General information 60 Basic Instructions for Use elisa 300 500 SW 2 09 x Rev 11 ...
Page 217: ...elisa 300 elisa 500 Instructions for Use Supplement A Manoeuvres Functions Ventilation Modes ...
Page 351: ...CO2 Mainstream Sensor LeoCap Instructions for Use Supplement B ...
Page 379: ...Multi Gas Sensor LeoLyzer Instructions for Use Supplement C ...
Page 407: ...Pulse Oximetry and LeoClac Instructions for Use Supplement D ...
Page 437: ...Medical Equipment Interface elisa megs Instructions for Use Supplement F ...
Page 440: ...4 Instructions for Use Supplement F elisa megs Rev 03 This page is intentionally left blank ...