Instructions for Use Supplement B – CO
2
Sensor LeoCap
|
Rev. 04
9
Warning
Sensor functions may be affected by equipment operated in the
vicinity. Before use, check that the device operates normally when
used in the vicinity of other equipment.
Warning
Keep magnetic objects away from the device. Magnetic objects
may distort the waveform and cause incorrect readings.
Warning
Do not use the device in the vicinity of nuclear magnetic reso-
nance equipment.
Warning
Measurements may not be performed correctly until several min-
utes after connection of the sensor because the temperature of
the device is not yet stable.
Warning
Do not allow liquid to enter into the housing of the device during
operation.
Warning
Do not immerse the device in liquids, such as water or disinfectant.
Warning
The CO
2
sensor is not defibrillation-proof. It is mandatory to re
-
move the CO
2
sensor from the patient during defibrillation.
Warning
Use only the specified airway adapters. Otherwise, the maximum
performance cannot be guaranteed due to larger dead space, leak-
age or an unsafe (electrical) connection, etc.
Warning
Do not use the adult/paediatric adapter on neonates because the
dead space of the airway adapter is 5 ml.
Summary of Contents for elisa 300
Page 3: ...elisa 300 elisa 500 Instructions for Use Basic Manual ...
Page 62: ...5 General information 60 Basic Instructions for Use elisa 300 500 SW 2 09 x Rev 11 ...
Page 217: ...elisa 300 elisa 500 Instructions for Use Supplement A Manoeuvres Functions Ventilation Modes ...
Page 351: ...CO2 Mainstream Sensor LeoCap Instructions for Use Supplement B ...
Page 379: ...Multi Gas Sensor LeoLyzer Instructions for Use Supplement C ...
Page 407: ...Pulse Oximetry and LeoClac Instructions for Use Supplement D ...
Page 437: ...Medical Equipment Interface elisa megs Instructions for Use Supplement F ...
Page 440: ...4 Instructions for Use Supplement F elisa megs Rev 03 This page is intentionally left blank ...