9 | Alarms and troubleshooting
132
Basic Instructions for Use elisa 300-500
|
SW 2.09.x
|
Rev. 11
Code Prior-
ity
Alarm
message
Cause
Corrective measure
#230
3
O2 flow
error
O2 flow
sensor error
Information:
- A redundant sensor is defective
- The accuracy of low readings can be affected
Ventilator is operating safely.
7
Information:
- The O2 flow sensor system is defective.
- Ventilation is now only possible with air
Monitor:
- O2 concentration
Decide: If ventilation with low oxygen is a hazard:
- Disconnect the patient from this ventilator
- Put this ventilator out of service
- Continue ventilation with another ventilator
- Contact an authorised service agent.
#231
3
Air flow
error
Air flow
sensor error
Information:
- A redundant sensor is defective
- The accuracy of low readings can be affected
Ventilator is operating safely.
7
Information:
- The air flow sensor system is defective
- Ventilation is now only possible with oxygen
Decide: If pure oxygen concentration is a hazard:
- Disconnect the patient from this ventilator
- Put this ventilator out of service
- Continue ventilation with another ventilator
- Contact an authorised service agent.
#233
3
O2 flow
error
O2 flow
sensor error
Information:
- A redundant sensor is defective
- The oxygen flow is limited to 80 l/min
Ventilator is operating safely.
7
Information:
- The O2 flow sensor system is defective.
- Ventilation is now only possible with air
Monitor:
- O2 concentration
Decide: If ventilation with low oxygen is a hazard:
- Disconnect the patient from this ventilator
- Put this ventilator out of service
- Continue ventilation with another ventilator
- Contact an authorised service agent.
Summary of Contents for elisa 300
Page 3: ...elisa 300 elisa 500 Instructions for Use Basic Manual ...
Page 62: ...5 General information 60 Basic Instructions for Use elisa 300 500 SW 2 09 x Rev 11 ...
Page 217: ...elisa 300 elisa 500 Instructions for Use Supplement A Manoeuvres Functions Ventilation Modes ...
Page 351: ...CO2 Mainstream Sensor LeoCap Instructions for Use Supplement B ...
Page 379: ...Multi Gas Sensor LeoLyzer Instructions for Use Supplement C ...
Page 407: ...Pulse Oximetry and LeoClac Instructions for Use Supplement D ...
Page 437: ...Medical Equipment Interface elisa megs Instructions for Use Supplement F ...
Page 440: ...4 Instructions for Use Supplement F elisa megs Rev 03 This page is intentionally left blank ...