Functions | A 3
Instructions for Use elisa 300-500 | Supplement A |
SW 2.09.x
| Rev. 11
79
A 3.13.6
Checklist for verification of the entire elisa 300 / elisa
500 system with the AnaConDa system
Check point:
Yes
No
System:
System test successfully completed according to chapter 6.5?
AnaConDa system securely connected between ventilation tube /
tracheostomy tube and Y-piece of the breathing circuit?
Anaesthetic gas filter or anaesthetic gas scavenging system connected
to expiratory flow sensor via adapter?
All tube connections checked for leaks and all plug connections veri-
fied?
Anaesthetic gas monitor connected via sidestream port and ready for
operation?
AnaConDa system filled according to manufacturer's instructions and
prepared for use?
Syringe pump ready for operation and properly connected?
Is a bag valve mask available for emergency ventilation?
Can the resuscitation bag be connected to the patient's tube / trache-
ostomy tube?
elisa 300 / elisa 500 device:
Is the internal battery fully charged?
Correct AnaConDa system (AnaConDa or AnaConDa-S) selected?
Activate the AnaConDa function and perform sedation with AnaConDa
(or AnaConDa-S) according to the manufacturer's instructions and
clinical circumstances.
Summary of Contents for elisa 300
Page 3: ...elisa 300 elisa 500 Instructions for Use Basic Manual ...
Page 62: ...5 General information 60 Basic Instructions for Use elisa 300 500 SW 2 09 x Rev 11 ...
Page 217: ...elisa 300 elisa 500 Instructions for Use Supplement A Manoeuvres Functions Ventilation Modes ...
Page 351: ...CO2 Mainstream Sensor LeoCap Instructions for Use Supplement B ...
Page 379: ...Multi Gas Sensor LeoLyzer Instructions for Use Supplement C ...
Page 407: ...Pulse Oximetry and LeoClac Instructions for Use Supplement D ...
Page 437: ...Medical Equipment Interface elisa megs Instructions for Use Supplement F ...
Page 440: ...4 Instructions for Use Supplement F elisa megs Rev 03 This page is intentionally left blank ...