Instructions for Use Supplement F – elisa@megs
|
Rev. 03
17
Electromagnetic immunity:
elisa@megs is intended for use in the electromagnetic environment specified
below. It is the responsibility of the customer or user to ensure that elisa@megs
is operated in such an environment.
Immunity
test
IEC 60601
test level
Compliance
level
Electromagnetic
environment –
guidance
Electrostatic dis-
charge
(ESD) to
IEC 61000-4-2
± 8 kV
contact
± 15 kV
air
± 8 kV
contact
± 15 kV
air
Floors should be wood,
concrete or ceram-
ic tile. If floors are
covered with synthetic
material, the relative
humidity should be at
least 30%.
Electrical fast
transients/bursts to
IEC 61000-4-4
± 2 kV for
power supply lines
± 1 kV for input/out-
put lines
± 2 kV for
power supply lines
± 1 kV for input/out-
put lines
The mains power quali-
ty should be that of a
typical commercial or
hospital
environment.
Surges to IEC
61000-4-5
± 1 kV
differential mode
± 2 kV common mode
± 1 kV
differential mode
± 2 kV common mode
The mains power
quality should be that of
a typical commercial or
hospital environment.
Voltage dips, short
interruptions and
voltage variations to
IEC 61000-4-11
100% dip in U
T
for
10 ms
100% dip in U
T
for
20 ms
30% dip in U
T
for
500 ms
100% dip in U
T
for
5000 ms
100% dip in U
T
for
10 ms
100% dip in U
T
for
20 ms
30% dip in U
T
for
500 ms
100% dip in U
T
for
5000 ms
The mains power
quality should be that of
a typical commercial or
hospital environment.
If the user of elisa@
megs requires contin-
ued operation during
power mains interrup-
tions, it is recommend-
ed that elisa@megs
be powered from an
uninterruptible power
supply (UPS).
Power frequency (50
Hz) magnetic field to
IEC 61000-4-8
30 A/m
30 A/m
Power frequency mag-
netic fields should be
at levels characteristic
of a typical location in
a typical commercial or
hospital environment.
Note: U
T
is the AC mains voltage prior to application of the test levels.
Summary of Contents for elisa 300
Page 3: ...elisa 300 elisa 500 Instructions for Use Basic Manual ...
Page 62: ...5 General information 60 Basic Instructions for Use elisa 300 500 SW 2 09 x Rev 11 ...
Page 217: ...elisa 300 elisa 500 Instructions for Use Supplement A Manoeuvres Functions Ventilation Modes ...
Page 351: ...CO2 Mainstream Sensor LeoCap Instructions for Use Supplement B ...
Page 379: ...Multi Gas Sensor LeoLyzer Instructions for Use Supplement C ...
Page 407: ...Pulse Oximetry and LeoClac Instructions for Use Supplement D ...
Page 437: ...Medical Equipment Interface elisa megs Instructions for Use Supplement F ...
Page 440: ...4 Instructions for Use Supplement F elisa megs Rev 03 This page is intentionally left blank ...