7 | Operation
84
Basic Instructions for Use elisa 300-500
|
SW 2.09.x
|
Rev. 11
7.2
Selection “Same Patient” / “New Patient”
A
B
By selecting “Same Patient” (
A
), you continue with the previously selected venti
-
lation mode including all ventilation parameters and alarm limits.
If “New Patient” (
B
) is selected, the alarm limits, ventilation mode and the ven
-
tilation parameters are all reset to the elisa 300 / elisa 500 default values. The
automatically activated default ventilation mode is the first mode on the tab bar.
Remark
If the default parameters are adapted to the patient condition pri-
or to starting ventilation, these will be reset to the defaults when
“New Patient” (
B
) is selected.
Remark
After selection of the “New patient”, the alarm tone volume will
be set to the level adjusted at the configuration level. The user
needs to make sure that the set alarm tone volume is adequate
for the clinical setting.
Summary of Contents for elisa 300
Page 3: ...elisa 300 elisa 500 Instructions for Use Basic Manual ...
Page 62: ...5 General information 60 Basic Instructions for Use elisa 300 500 SW 2 09 x Rev 11 ...
Page 217: ...elisa 300 elisa 500 Instructions for Use Supplement A Manoeuvres Functions Ventilation Modes ...
Page 351: ...CO2 Mainstream Sensor LeoCap Instructions for Use Supplement B ...
Page 379: ...Multi Gas Sensor LeoLyzer Instructions for Use Supplement C ...
Page 407: ...Pulse Oximetry and LeoClac Instructions for Use Supplement D ...
Page 437: ...Medical Equipment Interface elisa megs Instructions for Use Supplement F ...
Page 440: ...4 Instructions for Use Supplement F elisa megs Rev 03 This page is intentionally left blank ...