A 3 | Functions
38
Instructions for Use elisa 300-500 | Supplement A |
SW 2.09.x
| Rev. 11
the static control mode needs to be selected and set.
For safety and to prevent that elisa is switched off inadvertently while the
Cuffscout function is active, the Standby mode is indicated in large letters and a
warning is displayed stating that the cuff pressure is activated.
Another standby view is available which displays the cuff pressure as a live graph.
Remark
The Cuffscout needs to be deactivated before you can switch off
the ventilator.
Remark
This function is only available for the user
if enabled on the config
-
uration level and preset for the standby view.
When this view is selected, the Cuffscout graph is also displayed on the night
screen (see chapter “7.4.5.2 Display brightness” in the Basic Instructions for Use).