3 | General safety instructions
24
Basic Instructions for Use elisa 300-500
|
SW 2.09.x
|
Rev. 11
Device alarms
Monitoring
Gas supply
Electrical supply
Sensors
Turbine
User interface
Device control system
Visual and audible alarms
External components (if available):
CO
2
sensor, multi-gas sensor, Cuffscout, etc.
All device alarms are evaluated during ventilation, in standby
mode and when the automatic patient detection function does
not detect a patient. An alarm is triggered when applicable.
Patient alarms
Monitoring
Airway pressures
Tidal volume / minute volume
Respiratory rates (overall/spontaneous)
Leakage
O
2
concentration
In combination with the appropriate sensors: gas concentration of
CO2, sevoflurane, isoflurane
Patient alarms are only evaluated during ventilation and when
the automatic patient detection function detects a patient. An
alarm is triggered when applicable.
System test
As a matter of principle, a system test must be carried out before the ventilator
is used.
Warning
Never carry out a system test if a patient is connected to the
device.
The system test ensures that all safety-related functions and components are
tested before the ventilator is used. The system test can be bypassed in the event
of an emergency.
If the system test was bypassed, the system cannot determine the correct values
of the currently used breathing circuit which are necessary for optimal compli-
ance and resistance compensation. In this case, the measurement values of the last
Summary of Contents for elisa 300
Page 3: ...elisa 300 elisa 500 Instructions for Use Basic Manual ...
Page 62: ...5 General information 60 Basic Instructions for Use elisa 300 500 SW 2 09 x Rev 11 ...
Page 217: ...elisa 300 elisa 500 Instructions for Use Supplement A Manoeuvres Functions Ventilation Modes ...
Page 351: ...CO2 Mainstream Sensor LeoCap Instructions for Use Supplement B ...
Page 379: ...Multi Gas Sensor LeoLyzer Instructions for Use Supplement C ...
Page 407: ...Pulse Oximetry and LeoClac Instructions for Use Supplement D ...
Page 437: ...Medical Equipment Interface elisa megs Instructions for Use Supplement F ...
Page 440: ...4 Instructions for Use Supplement F elisa megs Rev 03 This page is intentionally left blank ...