Page 89 Leica EM TIC 3X Operating Manual 07/17
Make sure gas inlet holes are not visible in the clearance hole, if visible turn ion
source cartridge ~120 °. Hint: Orient the gas inlet holes in line with the thread
holes of the holding spring while inserting the cartridge.
Lock parts with holding spring and screws.
Unscrew and remove locking screw and insert ion source housing in the
instrument.
To reinstall the ion source carry out the steps 5.6.2.6
in the reverse order.
Reset the operation hour counter of the ion source!