Page 71 Leica EM TIC 3X Operating Manual 07/17
4.4.2.6 Menu
By pushing the Menu button following the information are available:
1. actual gun parameter and
2. process time
3. guns operation time (by pushing the displayed value the operation time can be
set to zero
4. pressure of process gas (should be adjusted within the green area)
5. temperature (in conjunction with the cooling stage)
6. LN
2
level of the Dewar (in conjunction with the cooling stage)
7. Vacuum value
Purge:
After installing the instrument, replacing the Argon bottle, or in case of unstable guns,
the Argon lines and the valves should be purged. Purging can only be done when the
working vacuum is reached. Press the Start button to perform the purging process.
After pushing the button, purging will be done for one minute. You may interrupt the
purging process at any time by pushing the button again or when leaving the menu.