Page 50 Leica EM TIC 3X Operating Manual 07/17
Connection of the LN
2
hose to the instrument
Connect the LN
2
hose with its threaded connection to the instrument.
Remove the yellow protective cup from the lower end of the pump. Slowly lower the
pump into the Dewar filled up with LN
2
. Hold the pump for a while until the strong
boiling decreases and carefully lower the pump until it is completely immersed.
The clean Dewar vessel has to be filled with LN
2
according to the safety
precautions.
Connect the LN
2
hose with its threaded connection to the LN
2
pump. Open the
shutter by pushing the lever before connecting the hose.
Please refer to the operating manual of the standard stage for proper usage.