Page 11 Leica EM TIC 3X Operating Manual 07/17
2.3
Dimensions and weight
Dimensions of the system with without light microscope and sound suppressor:
Width
:
500 mm
Height
:
485 mm
Depth
:
738 mm
Weight
:
approx. 62 kg
Dimensions of the shipping container:
Width
:
650 mm
Height
:
820 mm
Depth
:
950 mm
The weight of the transport-ready system is about 90 kg.
Volume of LN
2
Dewar: 25 l
2.4
Electrical and gas requirements
Voltage:
100
– 230 VAC, 50/60 Hz
Mains supply fluctuations:
± 15% of the nominal voltage
Fuses:
4 AT
Working gas:
preferably 99.999 % (Ar 5.0)
Gas flow:
<1 sccm/ion source with automatic control
2.5
Energy consumption, energy conditions
Power consumption:
Maximum:
300 W
2.6
Sound emissions
The continuous sound level emitted by the Leica EM TIC 3X Triple Ion Beam Cutter
system is about 50 dB (A).
2.7
IP code of the electrical devices
The electrical devices are equipped with housings and covers, and are therefore
designed to provide
degree of protection IP 20
(contact protection: blocking finger
access; protection from foreign bodies: protected from foreign bodies > d = 12 mm,
no special water protection).
2.8
Environmental conditions and limits for operation and storage
Operating temperature
:
+15
°C to +30 °C
Storage temperature
:
+5 °C to +40 °C
Atmospheric humidity for operation and storage: max. 80 % RH; non-condensing
Protection from dust and aggressive media must be provided during operation and
storage.