Page 97 Leica EM TIC 3X Operating Manual 07/17
6.
Troubleshooting
6.1
Error messages
In case of an error a message about the error and remedy suggestion will be
displayed on the screen.
6.2
Malfunction
Malfunction
System does not start when switched on
Cause
The cable is not connected to the mains power supply
Remedy
Connect to the mains power supply
Cause
Fuse on the main switch is faulty
Remedy
Replace fuse
Malfunction
Turbo pump does not reach setpoint speed
Cause
Vacuum leak
Remedy
Check for leak, first at the door and cover flange
Check the gas line connection
Malfunction
Coaxial light can
’t be switched by the touch
screen
Cause
It is only possible when a process is running
Remedy
Switching without processing can only be done
directly on the coaxial light
Malfunction
Ion guns unstable
Cause 1
Contaminated gas or gas line
Remedy
Clean gas line, check gas quality (Ar 5.0)
Cause 3
Contamination in ion guns
Remedy
Clean ion gun
Cause 4
Cathodes burned out
Remedy
Replace cathodes and anode