38
Instruction Manual V 1.1 - 11/2002
9.
System and error messages
•
GLASS SENSOR DEF
: The transport mechanism for the coverslips
features a sensor that detects broken glass or an empty coverslip
magazine. A defect of this sensor is displayed.
Remedy
: Check the sensor for dirt and clean thoroughly, if neces-
sary; if the problem persists, a service technician must be called.
•
TS5025 FULL
: Three empty input racks are located on the output
ramp of the TS5025. Remove all slide racks from the ramp.
•
TS5025 PRESENT
: Appears for 1 s during the initialization phase after
activating the main power switch. It indicates that TS5025 was rec-
ognized by the electronics.
•
ST5020 CONTROL
: Indicates that the most recently processed input
rack was received by the ST5020.
•
TS5025 WAITING
: An input rack was transferred from ST5020 to
TS5025, but it could not be processed (due to user intervention, e.g.
venting the mountant bottle) and is now waiting in the transfer posi-
tion of TS5025.
Remedy
: Prepare the CV5030 for operation (push in the bath guide,
prime the mountant bottle); this erases the display and the rack is
processed.
There are several error messages of the form ERROR N MM, (
N
one digit
and
MM
two digits). These error messages are intended for the service
department. If such an error message appears, the user can restart the
unit (after removing all obstacles) by pressing the
START
key. Please
read the following pages for more information!
If the error occurs repeatedly, contact the technical service.
Summary of Contents for CV5030