27
Leica CV5030 - Robotic Coverslipper
The coverslip or slide is released.
•
The
LIFT
key raises the output racks to the extraction position.
Prior to reinitialization, all slides must be removed from the out-
put racks!
•
The automatic action of the bath operation is deactivated by press-
ing
STOP
. The bath must be manually pulled from the bath channel if
it needs to be removed. However, it is not necessary to remove the
bath prior to reinitialization.
•
By pressing
START
, the unit can now be reinitialized.
(Same state as that following the activation of the unit)
•
Interruption by pulling out the bath guide:
•
The unit finishes the coverslipping process of the current slide
and pushes it into the output rack.
•
The bath is moved to the front position.
The unit switches to
READY
mode.
5.8Process-based interruptions
Empty coverslip magazine
•
In quick mode (selectable via user MENU, see section 5.4):
-
The coverslip transfer arm moves to the resting position,
-
ERROR
mode: Continuous alarm, display shows “
CHECK COVER
SLIPS
”,
-
Acknowledge the audible alarm by pressing the
RESPOND
key,
-
Fill the coverslip magazine,
-
Continue the coverslipping process with
START
(after acknowl-
edging the alarm).