23
Leica CV5030 - Robotic Coverslipper
5.
Operating the CV5030
Operating mode for parameter set n (in this example,
1
);
XXXXXX: FAST or SECURE; SECURE: gripper waits with taking a slide
from the input rack until P&P has found a coverslip;
•
[START] enables SECURE
•
[PAUSE] enables FAST (i.e. SECURE disabled)
•
[RESPOND]
→
Menu B main screen
•
[STANDBY] quits MENUs
Sets the limit value for the CS LOW display (universal parameter, ap-
plies to all parameter sets);
Procedure: place coverslip stack with desired height in coverslip
magazine
Press
START
!
P&P now moves to coverslip magazine and lowers to the top cover-
slip before returning to the INIT position
→
Procedure complete, new limit value stored
•
[RESPOND]
→
MAGAZINE LIMIT main screen
Error message if no coverslip found
•
[RESPOND]
→
MAGAZINE LIMIT main screen
From MAGAZINE LIMIT main screen:
•
[RESPOND] or [PAUSE]
→
Menu B main screen
•
[STANDBY]
→
quits MENUs
Summary of Contents for CV5030