37
Leica CV5030 - Robotic Coverslipper
•
CHECK LIFT
: Is always displayed if an irregularity with the output
racks is discovered. This may occur in the following cases: i) no
more slot is available during the coverslipping process; ii) no output
rack is found after pressing
START
; iii) not all slides were removed
from the output racks after pressing
LIFT
or after all input racks were
full during the coverslipping process.
Remedy:
Replace full output racks with empty ones or insert empty
racks.
• ST5020 REQUEST
: ST5020 maintains a slide rack for CV5030. But
CV5030 is not capable of directly accepting (e.g. it may be busy pro-
cessing another input rack); the message is generally accompanied
by an additional message about the cause of the prevention. This
prevention must be removed for the transfer to be successful (addi-
tional request through ST5020 after 60 s).
• COVER SLIP BROKEN
: A coverslip at the coverslip transport arm is
broken. Discard the coverslip by pressing the
RELEASE COVERSLIP
key; the coverslipping process can then be continued by pressing
the
START
key.
•
REFILL COVER SLIPS
: The coverslip magazine is empty; fill the cover-
slip magazine and continue the coverslipping process with
START
.
•
STOPPED BY USER
: The
STOP
key was pressed. The unit is
reinitialized by pressing the
START
key.
•
VACUUM IN PROC
: At the conclusion of a coverslipping process (FIN-
ISHED) or after 5 minutes in
PAUSE
mode, the mountant bottle is emp-
tied to low vacuum. This message appears if a new coverslipping
process is to be started during the pumping process. As soon as the
pumping is finished, the process can be started as usual.
9.
System and error messages
Summary of Contents for CV5030