21
Leica CV5030 - Robotic Coverslipper
5.
Operating the CV5030
5.4
User MENU
Pressing [RESPOND] for 2s while in READY or PAUSE mode opens
the User MENU
Menu A for the current parameter set n (in this example,
1
); the keys
associated with the abbreviations in the second line open the ap-
propriate settings menu;
•
Use [START]/[PAUSE] to switch parameter sets
•
[RESPOND]
→
Menu B
•
[STANDBY] quits the menu
Setting of the mountant stroke for the parameter set n (in this exam-
ple,
1
); NN=40, 50, 55 or 60; the mountant volume is adjusted to suit
the coverslip length;
•
[START]/[PAUSE] switches to higher/lower values
•
[RESPOND]
→
MENU A main screen
•
[STANDBY] quits MENUs
Setting of the mountant volume for the parameter set n; (in this exam-
ple,
1
) N=1-9; the mountant volume is controlled via the dispenser
opening time; raising by one step results in an increase by 25%;
•
[START]/[PAUSE] switches to higher/lower values
•
[RESPOND]
→
MENU A main screen
•
[STANDBY] quits MENUs
Mountant type for the parameter set n (in this example,
1
); N: 1-5; the
parameter directly affects the bottle pressure; raising by one step
means a pressure increase of 25%;
•
[START]/[PAUSE] sets the value
•
[RESPOND]
→
Menu A main screen
•
[STANDBY] quits MENUs