17
Leica CV5030 - Robotic Coverslipper
5.1
Electrical connection
The unit is dimensioned for:
100 to 240 V AC; 100 W; 50/60 Hz.
5.2
Key designations and their functions
START
starts the coverslipping process from READY and PAUSE mode
(exception: if PRIME is shown on the display, PRIME must be carried out
first) and after removable errors (coverslip magazine is empty, coverslip
is broken). After pressing the
STOP
key (display shows “
STOPPED BY
USER”
), it causes a reinitialization of the unit (see section
4.3
).
Pressing PAUSE
interrupts the coverslipping process (
BUSY LED
lights
up) after the currently processed slide is placed in the output maga-
zine;
PAUSE LED
lights up. Press
START
to continue.
Device parameters can be changed.
STOP
interrupts the unit operation immediately;
ERROR LED
lights up in
red. Use
START
to reinitialize the unit.
Pressing
LIFT
raises the output rack to the extraction position (not avail-
able while the
BUSY LED
is lit).
After the output rack reaches the extraction position, all slides
must be removed before the process can be started (again). Non-
observance can lead to damage of the slide.
Pressing
RESPOND
acknowledges an audible alarm; in
READY
or
PAUSE
mode, this key is used to enter the menu (holding the key pressed for 2 s)
to set the coverslip parameters.
The
RELEASE COVER SLIP
key releases the suction devices on the cov-
erslip transfer arm (locked while the
BUSY LED
is lit), and the coverslip is
released.
5.
Operating the CV5030
Summary of Contents for CV5030
Page 2: ......