31
Leica CV5030 - Robotic Coverslipper
•
If CV5030 is activated but not ready for processing a new input rack,
the ST5020 is prompted for a delayed request:
•
If priming is required: “
PRIME”
” display and audible signal.
•
If the output magazines are full: “
CHECK LIFT
” display and audible
signal.
•
If the output slide ramp TS5025 is full: “
TS5025 FULL
” display and
audible signal.
•
If the bath guide is pulled out: “
CHECK BATH
“ display and audible
signal.
-
Racks loaded manually and bath not removed.
-
Bath forward, does not move back (sensor sees slide/gripper arm
in front of rack).
-
Bath sensor does not respond (bath not present).
6.3
Interruption by the user
PAUSE
has the same function as described in section 5.3 (as long as the
bath guide is not pulled out, the slide rack is removed by the TS5025 if the
coverslipping process is continued or concluded)
.
The
STOP
key interrupts the process in an emergency (see section 5.3)
and places the CV5030 in
ERROR
mode. The input rack can remain in the
bath during reinitialization. Be sure to remove all racks from the bath!
The remaining slides can be processed by pressing the
START
key; the
empty input rack must then be manually removed from the bath (using
the door and pulling out the bath guide).
•
Pulling out the bath guide also interrupts the process. However, the
TS5025 transfer station is subsequently initialized and no longer
available for the removal of the empty slide rack.