34
Instruction Manual V 1.1 - 11/2002
8.
Cleaning and maintenance
8.1
Daily cleaning
•
After extended non-operating times and generally always before
priming, the nozzle must be checked for dried residues of the
mountant and cleaned, if necessary.
•
The suction devices for lifting coverslips must be regularly checked
for damage and dirt accumulation. Lightly contaminated suction de-
vices can be cleaned with a cloth moistened with solvent. Damaged
and heavily contaminated suction devices must be replaced.
•
The skids of the feeder mechanism and the face of the capacitive
sensor between the suction devices must also be checked for con-
tamination during the suction check and cleaned, if necessary.
•
Mountant buildup must be removed regularly from the headed dow-
els of the slide transport mechanism.
•
The gripper must be regularly cleaned of mountant residues and
glass fragments.
•
Output racks must be cleaned of mountant residues before they are
inserted.
•
The area of the bath transport must be regularly checked for fallen
glass fragments, coverslips and mountant.
•
Before inserting coverslips into the coverslip magazine, it must be
cleaned of any glass fragments and glass dust.
•
The fume filter must be changed at regular intervals.