System and error messages
• PRIME
: Appears after activating the unit from standby mode and vent-
ing the mountant bottle at the display. As long as this message is dis-
played, all keys except
PRIME
,
VENT MOUNTANT
,
STOP
and
STANDBY
are locked.
After pressing the
PRIME
key, the screen is cleared and all functions
are available.
•
BATH NOT READY
: The bath guide is pulled out so that coverslipping
is currently not possible (appears upon request through ST5020 and
by pressing the
START
key); place the bath on the bath transport belt
and push the bath guide inside until it stops; at that point, all unit func-
tions are available again.
•
BATH NOT PRESENT
: It appears if the bath which is the first to be
moved completely to the rear is not recognized after pressing
START
. In this case, either no bath is inserted or it is blocked in its
movement, e. g. through a fallen piece of glass.
Remedy
: Insert the bath (or remove it, remove blocking objects and
reinsert it), push the bath guide back until it stops.
•
CHECK BATH
: After finishing a manual cycle (FINISHED), empty input
racks must be removed. The bath guide is pulled out—the sensor re-
sponds. If this is not recognized between the end of a coverslipping
process and the
START
of the next one (either manually by pressing
START
or through request from ST5020), then this message appears.
Remedy
: Manually remove the empty input rack.