8
Instruction Manual V 1.1 - 11/2002
Operating the unit
The unit may only be operated by trained laboratory personnel.
It may only be used according to its intended use and the instructions in this instruction
manual.
Avoid skin contact with solvents or coverslip media.
Ensure correct dosage of mounting media. Excessive media may spill from the specimen
slide onto the working surface and thereby obstruct the movement of parts of the unit (notes
on cleaning).
During operation, do not block motor-driven components through handling or objects. Dan-
ger of personal injury through broken glass!
Unit must be operated with vent, or fume tubing and activated carbon filter must be as-
sembled and connected properly. Warning against hazardous fire and poison through sol-
vent vapors.
Cleaning and maintenance
The unit may be opened for maintenance or repair work by authorized service technicians
only.
Switch the unit off and unplug it before each cleaning.
Do not use any solvent containing acetone, chloride or xylene for cleaning the surface.
While operating and cleaning the unit, do not let any liquid enter the unit.
While handling cleaning materials, observe the safety regulations of the manufacturer and
the lab regulations.
Dispose of used reagents while observing applicable local regulations and the disposal
regulations of your company/lab.
2.
Safety