20
Instruction Manual V 1.1 - 11/2002
If not enough coverslips are present at startup, this will not
be recognized until the first slide is coverslipped!
The factory setting for the limit value is approx. 40 coverslips.
[PAUSE] was pressed; this message appears while the coverslipping
of the current slide is completed.
CV5030 is paused; press [START] to continue the coverslipping pro-
cess with parameter set n (in this example,
1
).
Coverslip magazine is empty; coverslipping operation interrupted
(accompanied by continuous audible signal).
20s audible signal after completion of the process, FINISHED remains
on display until a key is pressed or the bath opened.
Bath not OK; either no bath found (vessel could not be initiated after
[START]) or bath guide not actuated since last FINISHED (occurs
when pressing [START] or when connecting to ST5020)
Bath guide open and [START] pressed or query from ST 5020 (in this
case: regular 3s warning sequence)
Output rack either full or not present
The gripper could not take the slide in three attempts (slide crooked
or following slide too close)
A rack received from the ST5020 is processed with parameter set n
(in this example,
1
)
5.
Operating the CV5030