15
Leica CV5030 - Robotic Coverslipper
4.2.1 Vent system
Replace the activated carbon filter every six
months or when solvent vapors exit the air
outlet at the rear of the unit.
If xylene is used in the bath, the acti-
vated carbon filter must be replaced
every month!
1. Release the screw of the service flap (
1
) on
the left side of the unit.
2. Now release the Allen screw (
2
) and push
the cover up.
3. Insert the filter and tighten the screws in
the reverse order.
Use the supplied tools for assembly.
1
2
4.2.2 Coverslip collection tray and suction
cups
4.
Initial setup
Fig. 5
Fig. 6
The coverslip collection tray is hung on the
right side on the slide ejector. It can be re-
moved for cleaning purposes.
Fig. 7
1
2
4.2.3 Changing the
needle
Insert the needle into
the needle retainer (
1
)
and turn one corner
c o u n t e r c l o c k w i s e
over the plate.