16
Instruction Manual V 1.1 - 11/2002
4.4
Refilling consumables
4.
Initial setup
•
Fill the bath with xylene*.
•
Fill the bottle (
3
) for the paused position of
the dispensing valve with xylene*.
•
Insert the output rack (
4
) into the lateral
guide (on the left side of the unit.
1
2
4
3
4.3.
Transport lock
•
Slowly fill the mountant bottle with mountant
(
1
) (max. 250 ml) to avoid air bubbles. It is es-
sential that the mountant bottle stands open
for several hours (overnight) in the vent.
•
Screw the lid of the mountant bottle firmly
onto the bottle.
•
Select the coverslip magazine (
2
) for the de-
sired coverslip type (22 mm or 24 mm wide).
•
Fill the coverslip magazine with coverslips
(length 40-60 mm) and insert it.
Use the glass bottles from the stan-
dard delivery or suitable GLASS con-
tainers – but never use PLASTIC con-
tainers.
Transport
lock for
coverslip
transfer module
Transport
lock for the
gripper
Fig. 9
Fig. 8
*
xylene substitute, depending on the mountant
Fig. 10
Replace damaged or deformed suction cups.
To replace a suction cup, remove it by pulling it
downward and insert a new one.
Fig. 11