22
Instruction Manual V 1.1 - 11/2002
5.
Operating the CV5030
Correction of the mountant stroke for the parameter set n (in this ex-
ample,
1
); NNN can be a value from -100 to 0 (in increments of 10),
with -100 being a reduction by 10 mm; the total volume of mountant
does not change, but is distributed over a shorter line from a different
starting point;
•
[START]/[PAUSE] switches to higher/lower values
•
[RESPOND]
→
MENU A main screen
•
[STANDBY] quits MENUs
Menu B for the parameter set n (in this example,
1
); the keys associat-
ed with the abbreviations in the second line open the appropriate
settings menus;
•
[START]/[PAUSE] changes parameter set
•
[RESPOND]
→
Menu A
•
[STANDBY]
→
quits MENUs
Coverslip position for parameter set n (in this example,
1
); the value is
relative, NNN from -30 to 30 (-3 mm to 3 mm) in individual steps; 0 cor-
responds to the value of 60 in version 1.04
•
[START]/[PAUSE] changes value (in steps of 1, i.e. 0.1 mm)
•
[RESPOND]
→
Menu B main screen
•
[STANDBY] quits MENUs
Setting of the mountant stroke position for the parameter set n (in this
example,
1
); the value is relative (thus CORR for CORRECTION), NNN
can be a value from -100 to 100, with a positive operational sign re-
sulting in a shift toward the label field. 0 corresponds to the earlier
value 800; the value represents 1/10 mm, i.e. -50 corresponds to
-5 mm;
•
[START]/[PAUSE] changes to higher/lower value (in steps of 10)
•
[RESPOND]
→
MENU B main screen
•
[STANDBY] quits MENUs