19
Leica CV5030 - Robotic Coverslipper
5.
Operating the CV5030
5.3
Program flow
Activate power switch
→
device enters STANDBY mode (READY LED
red)
•
if the transfer station is connected, the message TS 5025
PRESENT will be displayed for 1s
[STANDBY] for 2s
→
device initialized
Welcome message
After initialization: priming required; extended warning signal if incor-
rect key pressed (e.g. [START]); if ST5020 is connected, a regular 3s
warning sequence will be triggered as the CV5030 is not prepared to
accept a rack
Dispenser located at P&P, i.e. not in the prime position (after startup
or when [PRIME] is pressed; accompanied by characteristic audible
signal)
Appears after pressing [START] if the dispenser is in the prime posi-
tion (accompanied by characteristic audible signal); periodic alarm
when connected to ST5020
(n=1-4): device is ready, process with parameter set n can be
launched by pressing [START]
Device coverslips with parameter set
n
(in this example,
1
)
Appears in the second line when P&P grips below the preset level
when getting the cover glass. The message is not cleared until the
sensor recognizes a higher level and otherwise remains active when
READY, PAUSE, PROCESSING etc.