28
Instruction Manual V 1.1 - 11/2002
5.
Operating the CV5030
•
In secure mode (see section 5.4)
-
The processes are identical to those in quick mode, but since no
slide has been removed from the bath, only a short alarm se-
quence is triggered to indicate the error state.
Coverslip broken
•
The automatic discharge automatically recognizes a broken cover-
slip and discards it in the discharge tray at the slide transport mecha-
nism.
•
After receiving a new coverslip, the process is automatically contin-
ued.
Output rack full
-
The unit automatically switches to
PAUSE
mode after filling the
last slot.
-
An audible alarm indicates that the output rack must be emptied.
Display: “
CHECK LIFT
”.
-
After emptying the output racks, operation can continue analo-
gous to the regular
PAUSE
mode.
-
If the end of the input rack is reached at the same time,
CHECK
LIFT
is not displayed; instead, the process is concluded with
FIN-
ISHED
.
•
Error in the unit or process
-
ERROR
mode: Continuous alarm, display shows “
ERROR
” + error
number; the condition is identical to pressing
STOP
.
Summary of Contents for CV5030